Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 8000A: CAL ADJ /5520,6666 DATE: 18-Jan-95 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 99 CONFIGURATION: Fluke 5520A STANDARD: Philips PM 6666 ============================================================================= # # Source: # Fluke 8000A instruction manual (PN 347906). # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5520A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 STD Philips PM 6666 1.002 ASK- N J W 1.003 HEAD EQUIPMENT SETUP 1.004 DISP Remove the screw located just below the AC line 1.004 DISP jack. Slide the instrument out of the case. 1.005 DISP Connect the UUT to an ac power source. 1.005 DISP Turn the UUT on. 1.005 DISP Warm-up time: 1 hour. 1.005 DISP Ambient temperature: 22C - 25C. 1.005 DISP Relative humidity: less than 70%. 1.006 HEAD Period Adjustment 1.007 DISP Connect the PM 6666 and UUT as follows: 1.007 DISP [32] PM 6666 INPUT A HI to UUT TP5 1.007 DISP [32] PM 6666 INPUA A LO to UUT INPUT LO 1.008 DISP Select Period A on the PM 6666 front panel. 1.009 DISP Adjust R20 for a time period of 100ms (+/-5us). 1.009 DISP Variations of the time period must be <= 15us. 1.010 DISP Disconnect the PM 6666 from the UUT. 1.011 HEAD ZERO OFFSET ADJUSTMENT 1.012 DISP Connect the 5520A and the UUT as follows. 1.012 DISP [32] 5520A NORMAL HI to UUT V-Ohm 1.012 DISP [32] 5520A NORMAL LO to UUT COMMON 1.013 RNG 200mV 1.014 5520 0mV S 2W 1.015 DISP Adjust R15 for a UUT display of 00.0+/-00.1. 1.016 5520 * S 1.017 OPBR Does the UUT read less than or equal to 01.0? 1.018 JMPF 1.015 1.019 HEAD TURN_OVER ERROR ADJUSTMENT 1.020 5520 190mV S 2W 1.021 DISP Adjust R23 for a UUT reading of 190.0mV. 1.022 5520 -190mV S 2W 1.023 DISP Adjust R15 for a UUT reading of 190.0+/-0.1mV. 1.024 HEAD RANGE ADJUSTMENTS (DC Volts) 1.025 5520 190mV S 2W 1.026 DISP Adjust R23 for a UUT reading of 190.0+/-0.3mV. 1.027 5520 * S 1.028 RNG 2V 1.029 5520 1.9V S 2W 1.030 DISP Adjust R25 for a UUT reading of 1.900+/-0.003V. 1.031 5520 * S 1.032 HEAD RANGE ADJUSTMENTS (Ohms) 1.033 RNG 20MZ 1.034 5520 19MZ S 2W 1.035 DISP Adjust R55 for a UUT reading of 19.00+/-0.11 MOhm. 1.036 5520 * S 1.037 DISP Connect the 5520A and the UUT as follows. 1.037 DISP [32] 5520A NORMAL HI to UUT V-Ohm 1.037 DISP [32] 5520A NORMAL LO to UUT COMMON 1.037 DISP [32] 5520A AUX HI to UUT V-Ohm 1.037 DISP [32] 5520A AUX LO to UUT COMMON 1.038 RNG 20kZ 1.039 5520 19kZ S CW 1.040 DISP Adjust R12 for a UUT reading of 19.00+/-0.05 kOhm. 1.041 5520 * S 1.042 HEAD RANGE ADJUSTMENTS (High Freq. AC) 1.043 DISP Connect the 5520A and the UUT as follows. 1.043 DISP [32] 5520A NORMAL HI to UUT V-Ohm 1.043 DISP [32] 5520A NORMAL LO to UUT COMMON 1.044 RNG 20V AC 1.045 5520 19V 20kH SI S 2W 1.046 DISP Adjust C3 for a UUT reading of 19.00+/-0.21V. 1.048 END